Operation Manual touchTymp MI 26 and MI 36 Version
8511780 Rev. 4
79
30/06/2021
6 Technical Data
This section offers you important information about
▪
the touchTymp hardware specifications
▪
connections
▪
the pin assignment
▪
immittance and audiometry calibration values
▪
electromagnetic compatibility (EMC)
▪
electrical safety, EMC and associated standards
6.1 touchTymp Hardware
The touchTymp is an active, diagnostic medical
product according to the class IIa of the Medical
Device Regulation (EU) 2017/745.
General Information About Specifications
The performance and specifications of the device can only be guaranteed if it is
subject to technical maintenance at least once every 12 months.
MAICO Diagnostics puts diagrams and service manuals at the disposal of
authorized service companies.
STANDARDS
Medical CE-mark
Yes
Safety Standards
IEC 60601-1:2005+A1:2012/
ANSI/AAMI ES60601-1: 2005 / A2:2010/
CAN/CSA-C22.2 No. 60601-1:14
Type B Applied Parts
EMC Standards
IEC 60601-1-2
Audiometer
Standards
Tone: IEC 60645-1 Type 3/ANSI S3.6 Type 3
Tympanometer
Standards
IEC 60645-5, Type 2/ANSI S3.39, Type 2
DEVICE SPECIFICATIONS
Power supply
Type
UES65-240250SPA3
Input
90 to 264 V AC, 50/60 Hz, 2.0 A
Output
24.0 DC, 2.5 A MAX
Safety
IEC 60601-1, Class I
Mode of operation
Continuous