The device is not intended for operation in areas with an
explosion hazard. Do NOT use the device in a highly
oxygen-enriched environment, such as a hyperbaric
chamber, oxygen tent, etc. If the device is not used switch
it off and disconnect it from the power supply.
Never short-circuit the terminals.
To avoid the risk of electric shock, this equipment must
only be connected to the medical power supply originally
delivered by MAICO. Using another power supply can also
lead to electrical damage on the device.
Prevent cable breakage: cables must not be bent or
buckled.
2.7 Device Control
The user of the device should perform a subjective device check once a week
according ISO 8253-1. See section 6.7 for a checklist.
For annual calibration see section 2.5 and 3.2.
2.8 Electromagnetic Compatibility (EMC)
This device is suitable in hospital environments except for
near active HF surgical equipment and RF shielded rooms
of systems for magnetic resonance imaging, where the
intensity of electromagnetic disturbance is high.
The device fulfills the relevant EMC requirements. Avoid
unnecessary exposure to electromagnetic fields, e.g. from
mobile phones etc.
Use of this device adjacent to or stacked with other
equipment should be avoided because it could result in
improper operation. If such use is necessary, this device
and the other equipment should be observed to verify that
they are operating normally.
Use of accessories, transducers and cables other than
those specified or provided by the manufacturer of this
equipment could result in increased electromagnetic
emissions or decreased electromagnetic immunity of this
equipment and result in improper operation.
The list of accessories, transducers and cables can be
found in the section 6.5 of this instruction.
Portable
RF
communications
equipment
(including
peripherals such as antenna cables and external antennas)
should be used no closer than 30 cm (12 inches) to any part
of the touchTymp, including cables specified by the
manufacturer. Otherwise, degradation of the performance of
this equipment could result in improper operation.
WARNING
WARNING
WARNING
WARNING
WARNING
WARNING