5.6.14 Settings
– ETF – Perforated (MI 36 Version)
Figure 142
Figure 143
Start pressure:
the pressure that is first introduced when
performing
ETF
– Perforated
measurement (Figure
142). This is a default setting and can be configured
within the test screen.
NOTE:
You can adjust the pressure in 25 daPa steps.
Test duration
: Defines the length the time the test will be
conducted (Figure 143). Test duration can be configured
in 5 s increments from 30 s to 100 s.
5.6.15 Settings
– Audiometry – General
Figure 144
Signal
: Select a default signal for the Audiometry Module
Choose
Steady, Pulse, Warble
or
Pulse & Warble
Figure 145
Pulse length
: The length of each tone when Pulse is selected
(Figure 145). Choose between
250 ms
and
500 ms
.
Figure 146
Default hearing level
: Select a level the test signal will
return to with the following actions:
•
Enter Audiometry module
•
Change ear
•
Change transducer
•
Delete
•
•
Transfer to PC.
Choose between
-10 dB HL
to
50 dB HL
in
5 dB
steps
Figure 147
Level steps
: Select the change in volume with each + or -
selection of Hearing Level and Masking Level. Choose
between: 1 dB and 5 dB steps (Figure 147).
Figure 148
Test mode
: Selection of the operation mode of the
audiometer (Figure 148). Choose between:
•
Presenter
: Tone is presented when the
Present
icon is touched.
•
Interrupter
: Tone is interrupted/stopped when the
Present
Icon is touched.
Figure 149
Presenter duration
: The length of the signal with one
touch of the
Present
button. Choose between
200 ms,
300 ms, 400 ms, 500 ms, 700 ms, 1000 ms, 2000 ms,
3000 ms
or
Unlimited
(will present only when button is
touched) (Figure 149).
NOTE
: Setting is only relevant when
Presenter
is
selected for
Test mode
.