Operation Manual touchTymp MI 26 and MI 36 Version
8511780 Rev. 4
42
30/06/2021
Figure 49
Table:
The measured intensity level in dB HL is displayed
simultaneously to the ongoing test. Underneath each diagram
the frequency is displayed (Figure 49).
At the conclusion of the test, the result symbol appears at the top right corner of
the box either in the graphical view as in the table view. This is displayed for the
Acoustic Reflex
measurement that meets the criteria as defined in the setup
menu. A green checkmark indicates a present reflex. A red cross indicates
No Response
. To be considered as a
Pass
the maximum amplitude of the
reflex shape must reach a defined value (sensitivity) for a defined time.
Otherwise it is considered as
No Response
.
Noise stimuli (MI 36 Version Only)
Figure 50
The MI 36 version includes pure tone and noise stimuli for
Acoustic Reflex
testing.
:
Pure tone
(500 Hz, 1000 Hz, 2000 Hz, 4000 Hz)
:
Noise
(BB
– Broadband, HP – High Pass, LP – Low Pass)
Select the stimulus type to set or confirm test stimuli prior to
starting test. When the button is blue, this notes there is an
active stimulus to be tested. When both buttons are blue
at
least
one pure tone and one noise stimulus will be presented
during the test.
Edit Acoustic Reflex
Acoustic reflex results can be reviewed by the
Edit
button within the tool bar. When
this button is selected, the device is in edit mode where results can be reviewed or
modified prior to printing or software transfer (Figure 51). The edit mode is only
available when the display
Presentation
mode is set to
Graph
in the
Settings
(see
section 5.6.9).
NOTE
:
Edit
button is only available for selection when a result has been stored on
the screen.
Figure 51