Operation Manual touchTymp MI 24 and MI 34 Version
8107420 Rev. 6
45
03/09/2018
Figure 72
Compare the single tympanograms in the multilayered
tympanogram (Figure 72). Tympanograms displayed
include:
Red
or
Blue
: represents test ear
Orange
:
represents “
Swallow
”
Green:
represents “
Valsalva maneuver
”
NOTE:
The order of instructions displayed can be
configured in the
Settings
Performing an ETF Perforated Measurement
Figure 73
ETF Perforated
determines if the patient can open
his/her Eustachian tube when the TM is perforated or an
open PE-tube is in place.
ETF Perforated
will put the
middle ear under a certain
Start pressure
based on the
default setting, but can be modified in 25 daPa steps by
the on screen pressure setting (Figure 73).
Figure 74
The graph displays the vertical axis as pressure, and the
horizontal axis as time (Figure 74).
Instruct the patient not to move or talk until the test is over.
When a seal is obtained, the device displays a message
to swallow as many times during the test duration.
NOTE
: Automatic and manual mode are operated the same
for this
ETF Perforated
test as a start operation is required.
Figure 75
Pressure will increase to the predetermined setting.
Let the pressure run a few seconds at peak pressure to
verify a successful seal. Once the peak pressure has
been obtained ask the patient to swallow as many times
as they can while the test is running.
If the
Eustachian tube
opens, a drop in pressure will be
recorded. Repeated attempts to swallow will display a
downward stair step effect, or a complete drop to 0 daPa
(Figure 75).
Numerical results of the test are displayed below the
graph. Each time the device detects opening and closing
of the
Eustachian tube
, the results are recorded. An
open and close result is displayed up to three values.
The test will stop after the allotted time (i.e. 30 seconds)
as defined in the settings or the examiner manually
stops the test.