Operation Manual touchTymp MI 24 and MI 34 Version
8107420 Rev. 6
17
03/09/2018
Figure 7
Screw the probe cap back on the probe (Figure 7). The
force of tightening the cap will tighten the screw
sufficiently. Never use tools to fix the probe cap!
If any blockage or damage occurs to the sealing gasket,
the probe system can only be serviced by MAICO.
Cleaning alternative
Figure 8
Figure 9
Use the cleaning set from the eartip box (Figure 8): Take
the cleaning tool apart to find the thin brush and thin rigid
plastic cord (Figure 9).
Figure 10
Use the plastic cord or brush to push debris out of the
probe tip (Figure 10).
Figure 11
Always enter the probe tip from the rear to avoid
accumulation of debris inside the vents (Figure 11).
Figure 12
Never clean the probe itself with
the cleaning devices. The probe
will be damaged (Figure 12).
Figure 13
Never clean the probe tip while
the tip is still attached. The probe
will be damaged (Figure 13).
3.4 Disposables
Figure 14
Operating the touchTymp will require the use of eartips
–
either mushroom shaped (1) or umbrella (2) eartips
(Figure 14).
Eartips are intended for single-use only. These should be
discarded after use. They cannot be cleaned.
CAUTION
CAUTION
1
2