Operation Manual touchTymp MI 24 and MI 34 Version
8107420 Rev. 6
13
03/09/2018
In order to maintain a high level of safety and to ensure the
device works properly, it is necessary to have the device
and its power supply checked according to the medical
electrical safety standard IEC 60601-1 by a qualified
service technician at least once a year. For more
information see section 3.2.
The use of non-calibrated devices can lead to incorrect test
results and is not advisable.
Prevent cable breakage: cables must not be bent or buckled.
2.7 Device Control
The user of the device should perform a subjective device check once a week
according ISO 8253-1.
For annual calibration see section 2.5 and 3.2.
2.8 Electromagnetic Compatibility (EMC)
Electrostatic discharge (ESD) according to IEC 61000-4-2.
Use the device only in an electrostatic controlled
environment.
To avoid the risk of electric shock, this equipment must
only be connected to supply mains with protective earth.
The device fulfills the relevant EMC requirements. Avoid
unnecessary exposure to electromagnetic fields, e.g. from
mobile phones etc. If the device is used adjacent to other
equipment it must be observed that no mutual disturbance
appears.
Please also refer to EMC consideration in section 6.5.
CAUTION
WARNING