Operation Manual touchTymp MI 24 and MI 34 Version
8107420 Rev. 6
44
03/09/2018
5.4.4.5 Performance and Evaluation of Eustachian Tube Function (ETF)
– MI 34 Version
Selection of the
ETF
icon from the
Home
screen or
Fixed Function Bar
leads to the
ETF
ETF
has two operations:
ETF Intact
: performed on patients with normal tympanic membrane (TM).
ETF Perforated
: performed on patients with a perforated TM or open PE tubes
in place.
ETF Intact
is the default selection when the module is entered.
Figure 67
Performing a Measurement
Figure 68
Select the test type
ETF Intact
, or
ETF Perforated
Performing an ETF Intact Measurement
ETF Intact
is performed by measuring three tympanograms on a multilayer display.
Before testing begins instruct the patient not to move or talk until the test is completed.
Any sound or movement may give unreliable results.
Figure 69
As the test is progressing the numerical information
below the graph is displayed. Once the first
tympanogram
is complete, the pressure at the
maximum compliance appears under
Pressure 1
Figure 70
The pressure is held as the patient is instructed to
perform a maneuver (i.e.
Swallow
,
Valsalva
) (Figure
70). When completed, press continue for the second
tympanogram
to be completed. The pressure at the
maximum compliance appears under
Pressure 2
.
Figure 71
Once again the pressure is held while the instruction is
displayed for the patient to perform the second
maneuver (Figure 71). Press continue to perform the
third
tympanogram
. The pressure at the maximum
compliance displays under
Pressure 3
.