5.4.3.5 Status Indicator
Figure 40
The status indicator (Figure 40) in the middle of each
test screen provides the probe status on the display
screen.
The same information is shown on the probe with the single LED (Table 11).
Table 11 Test Status Indication
TEST STATUS INDICATION
PROBE
SCREEN
INFORMATION
Right ear is selected. Probe is out of ear.
Left ear is selected. Probe is out of ear.
Probe is in the ear and is sealing, test is running or
test is done.
Probe is in the ear and blocked or leaking.
1. Reinsert probe for better placement.
2. Check eartip size and condition.
3. Inspect probe tip for any blockage.
No Light
Probe is not attached properly. Check probe
connection.
No Light
Probe tone is not given. This status is shortly
shown while the frequency is being changed.
5.4.4 Testing
5.4.4.1 Performance and Evaluation of Tympanometry Test
Figure 41 shows the
Tympanometry
test screen.
Figure 41
NOTE
: Tympanometry test screen explanation apply to the
Tympanometry
module and
the
Tympanometry and Acoustic Reflex
module.