![Maico MI 24 Operation Manual Download Page 13](http://html2.mh-extra.com/html/maico/mi-24/mi-24_operation-manual_3681496013.webp)
Operation Manual touchTymp MI 24 and MI 34 Version
8107420 Rev. 6
12
03/09/2018
In Case of Emergency
In case of emergency, disconnect the device from power supply.
Do not position the device in a way that it is difficult to
operate the disconnection device. The supply mains and
the power socket shall be accessible at all times.
Do not use the device if the mains cable and/or the plug is
damaged.
To transfer data to a PC, establishing a PC-connection via
USB is required. See section 4.2.4 on how to safely establish
a connection with a power supplied PC or laptop (medical
device/non-medical device) or to a battery-driven laptop.
This equipment is intended to be connected to other
equipment thus forming a Medical Electrical System. External
equipment intended for connection to signal input, signal
output or other connectors shall comply with the relevant
product standard e.g. IEC 60950-1 for IT equipment and the
IEC 60601-series for medical electrical equipment. In addition,
all such combinations
– Medical Electrical Systems – shall
comply with the safety requirements stated the general
standard IEC 60601-1, edition 3, clause 16. Any equipment
not complying with the leakage current requirements in IEC
60601-1 shall be kept outside the patient environment i.e. at
least 1.5 m from the patient support or shall be supplied via a
separation transformer to reduce the leakage currents. Any
person who connects external equipment to signal input,
signal output or other connectors has formed a Medical
Electrical System and is therefore responsible for the system
to comply with the requirements. If in doubt, contact qualified
medical technician or your local representative. If the device
is connect to a PC (IT equipment forming a system) ensure
not to touch the patient while operating the PC.
If the device is connect to a PC (IT equipment forming a
system) assembly and modifications shall be evaluated by
qualified
medical
technician
according
to
safety
regulations in IEC 60601.
The device is not intended for operation in areas with an
explosion hazard. Do NOT use the device in a highly
oxygen-enriched environment, such as a hyperbaric
chamber, oxygen tent, etc. If the device is not used switch
it off and disconnect it from the power supply.
Never short-circuit the terminals
To avoid the risk of electric shock, this equipment must
only be connected to the medical power supply originally
delivered by MAICO. Using another power supply can also
lead to electrical damage on the device.
WARNING
WARNING
WARNING
WARNING
Summary of Contents for MI 24
Page 1: ...Operation Manual touchTymp MI 24 and MI 34 Version...
Page 2: ......