Operation Manual touchTymp MI 24 and MI 34 Version
8107420 Rev. 6
55
03/09/2018
Figure 122
for MI 34 version also include noise stimuli (
BB
–
Broadband,
HP
– High Pass,
LP
– Low Pass).
When options are greyed out in the settings screen, the
license is not active (Figure 122).
5.6.12 Settings
– Decay – General (MI 34 Version)
Figure 123
Figure 124
Duration:
Defines the length the tone will be presented
to the patient (Figure 123).
Duration
can be configured
in by 5 second increments from 10 s to 30 s.
Level
: Defines the default intensity of the stimulus upon
entering the screen (Figure 124). You can adjust the
level in 5 dB steps.
NOTE:
The deflection of the acoustic reflex is defined
within the
Acoustic Reflex
settings.
5.6.13 Settings
– ETF – Intact (MI 34 Version)
Figure 125
First Test:
Defines the message while the test is in
progression. The end user can select which maneuver
will be displayed first,
Swallow or Valsalva
NOTE:
The selection also determines the color of the
tympanometry graphs:
Swallow
represented by
Orange
Valsalva
represented by
Green
Figure 126
Pump speed
: Selection of pump speed determines how
precisely and quickly the test will proceed (Figure 126).
NOTE:
A slow speed is more time consuming, but
may give more detailed information.
There are four different
pump speed
Settings:
Automatic
(Dynamic from 600 daPa/s for low gradient
and 200 daPa/s for a gradient larger than 5 daPa)
Minimum
(50 daPa/s): slow, very precise results
Medium
(250 daPa/s): compromise of speed and precision
Maximum
(>400 daPa/s): fast, screening