73
47DT OPERATING MANUAL EM-9508-B Rev.7
■
TEMPERATURE UNIT: °F
MAIN DISPLAY
FUNCTION
USABLE RANGE
CONFORMANCE RANGE
DEFAULT VALUE
TYPEK1
K (CA)
-382.0 to +2426.0
-328.0 to +2372.0
Analog output 0%:
Analog output 100%:
TYPEK2
K (CA), narrow span
-58.0 to +986.0
-4.0 to +932.0
Analog output 0%:
Analog output 100%:
TYPE E
E (CRC)
-22.0 to +1166.0
32.0 to 1112.0
Analog output 0%:
Analog output 100%:
TYPEJ1
J (IC)
-202.0 to +1616.0
-148.0 to +1562.0
Analog output 0%:
Analog output 100%:
TYPEJ2
J (IC), narrow span
-58.0 to +806.0
-4.0 to +752.0
Analog output 0%:
Analog output 100%:
TYPE T
T (CC)
-382.0 to +806.0
-328.0 to +752.0
Analog output 0%:
Analog output 100%:
TYPE B
B (RH)
158.0 to 3326.0
752.0 to 3272.0
Analog output 0%:
Analog output 100%:
TYPE R
R
-22.0 to +3146.0
32.0 to 3092.0
Analog output 0%:
Analog output 100%:
TYPE S
S
-22.0 to +3146.0
32.0 to 3092.0
Analog output 0%:
Analog output 100%:
TYPE C
C (WRe 5-26)
-22.0 to +4226.0
32.0 to 4192.0
Analog output 0%:
Analog output 100%:
TYPE N
N
-382.0 to +2426.0
-328.0 to +2372.0
Analog output 0%:
Analog output 100%:
TYPE U
U
-382.0 to +806.0
-328.0 to +752.0
Analog output 0%:
Analog output 100%:
TYPE L
L
-202.0 to +1616.0
-148.0 to +1562.0
Analog output 0%:
Analog output 100%:
TYPEPR
(PR)
-22.0 to +3146.0
32.0 to 3092.0
Analog output 0%:
Analog output 100%:
IMPORTANT
• The operational range of the DC output is -10 to +110% of the output span after analog output 0% and 100% adjust-
ments.
• The output is saturated at approximately -10% or +110%, or [max. value of the usable range + 20°C (+ 36°F)] or [min.
value - 20°C (- 36°F)].
• The analog outputs 0% and 100% are returned to the previously set values per input type (or default values when the
input type is selected for the first time) when the input type has been changed.
• The analog outputs 0% and 100% can be set to any value, however it is recommended to use them within the usable
range.