52
47DT OPERATING MANUAL EM-9508-B Rev.7
3
Press Alarm/
↓
or Init/
↑
button to go to the decimal point posi-
tion setting.
• ‘0.0’ is indicated.
• The SD indicates ‘D-POINT’.
NOTE
Skip to Step 7 if the default value is acceptable.
4
Press Shift button to shift the display into the setting standby
mode.
• The indication starts blinking, to which you can apply changes.
5
Press Up button to select ‘0’.
6
Press Alarm/
↓
or Init/
↑
button to apply the new setting.
• And the next parameter setting is indicated.
NOTE
• Press Alarm/
↓
button, and the input compensation A input value will be indicated (‘CR-A IN’ on the SD).
• Press Init/
↑
button, and the temperature unit will be indicated (‘UNIT’ on the SD).
7
■
TO GO ON TO SET THE BARGRAPH TYPE,
Skip to Step 2 in “8.1 BARGRAPH TYPE”.
■
TO QUIT,
Hold down Alarm/
↓
or Init/
↑
button for 1 second or more to return to Measuring Mode.
Blinking
Blinking