41
8.1.2. Automatic measurement mode
!
NOTE
The automatic measurement mode is composed of optimum measurement conditions and can
be measured reliably. If the patient's eyes move and it is difficult to measure, press the
measurement button on the joystick.
The mode gets converted to automatic measurement mode when MANUAL button is pressed on while in
the manual measurement mode.
While in the automatic measurement mode, measurement is carried out automatically even when the
measurement button is not pressed on when the state reaches a state in which arrangement in the device
and measurement is realized effectively.
①
(Eye height adjustment), (measurement location and focusing) process is carried out just like the
manual measurement mode.
②
Measurement
-
Measurement is carried out automatically when the location arrangement and focusing are
completed.
-
Value for new measurement result appears on the monitor screen after measurement takes place
up to the frequency (possible to select among three, five and continuous) designated on the user
Setup mode.
-
Up to 99 measurement frequencies are indicated and it is possible to check the measurement
values
up to the latest 10 times once again in the Display mode.