NOTE
8.2.2. Automatic measurement mode
The automatic measurement mode is composed of optimum measurement conditions and can
be measured reliably. If the patient's eyes move and it is difficult to measure, press the
measurement button on the joystick.
The mode gets converted to automatic measurement mode when MANUAL button is pressed on while
in the manual measurement mode. In case of automatic measurement mode, when the state reaches a
state in which the arrangement in the device and measurement is realized effectively, measurement takes
place automatically even when the measurement button is not pressed on.
①
Location arrangement and focus are adjusted just like the (measurement location and focusing) while
in the 8.1.2 refractive power measurement mode.
②
Measurement takes place automatically using the same method as that of the (measurement)
process while in the 8.1.2 refractive power measurement mode.
③
Measurement result is printed using a method that is same as that of the (printing) process while in
the 8.1.2 refractive power measurement mode.
8.2.2. Message List
"MOVE RIGHT"
Move stage to right.
"MOVE LEFT"
Move stage to left.
"CHINREST DOWN"
Move chinrest down.
"CHINREST UP"
Move chinrest up.
"DATA
TRANSMITTING"
Measurement data of LRK is being transmitted to external.
"DATA PRINTING"
Measurement data of LRK is being printed.
"HLM DATA
PRINTING"
Measurement data of LM is being printed.
"HDR DATA
TRANSMITTING"
Measurement data of LM is being transmitted to external.
"FINISH"
(Auto shot mode) Measurement is finished.
“ALIGN ERROR”
The alignment (focus or center) is significantly failed during the
measurement.
“NO SIGNAL”
- Center or eye cannot be found.
- The patient’s eye blinks or moves during measurement.
- If this message appears while measuring model eye, the
instrument
may have a problem. Contact your service engineer.
"TRY AGAIN"
There is too big difference from the previous measurement value.