22
The section “Default SLE
D Drive Current
–
PC Mode” all
ows user to set the initial current
levels for each SLED when the unit is turned on and is in PC MODE.
The section “Default SLED Drive Current –
MANUA
L MODE” consists of the operati
ng
current at which each SLED will be driven when the ISB2 is operated in MANUAL MODE.
Note: SLEDs should be operated per manufacturer-recommended current setting
provided in the Final Test Inspection Report for the BeST-SLED® ISB2. Running the SLEDs
at higher currents may cause damage. Luxmux is not liable for damage if SLEDs are
operated at higher than specified current ratings.
The
section “Default TEC Temperature” allows user
s to change the default settings for
the TEC temperature set point on the SLED TEC.
Changes and notifications are displayed in the text box in the lower-right quadrant of
the Set Defaults window.