Back Panel - Thermal Management:
Free air circulation around the rear of the unit should be maintained for good thermal
performance. If the ISB2 is operated in environments over 35 Degrees Celsius the ISB2
should be mounted to an external heatsink. Free air circulation around the ISB2 is
required when it is used without a heatsink. A minimum of 0.5
”
clearance is
recommended for free air circulation around the top and sides of the ISB2 Unit.
Air circulation is located on the back panel of the ISB2 Light Source, while additional
ventilation holes are located in the front panel. Ensure vents are not obstructed to allow
for maximum air flow.
The back panel includes two fans as shown in Figure 4. Upon powering up the unit, each
fan is regulated automatically to its maximum speed. Users do have the option,
however, to control fan speed manually as explained in the Software Operation section.
If the fan is not working and the heat sink temperature goes over 60
˚
C, turn the unit off
and contact Luxmux technical support.
Figure 4: Back Panel
–
Thermal Management
Summary of Contents for BeST-SLED ISB2