- 53 -
Electrical fast
transients /
bursts
(IEC 61000-4-4)
Power lines: 2 kV;
100 kHz repetition
frequency
Signal lines: 1 kV;
100 kHz repetition
frequency
Mains power quality should be that of a
typical environment.
Surges
(IEC 61000-4-5)
L-N: 1kV at 0°, 90°,
180°, 270°
L-PE, N-PE: 2 kV at
0°, 90°, 180°, 270°
Mains power quality should be that of a
typical environment.
Conducted
disturbances
inducted by RF
fi elds
(IEC 61000-4-6)
0.15-80 MHz; 1kHz
AM 80 %; 3 Vrms,
6 Vrms in ISM and
amateur radio
band
Portable and mobile RF communica-
tions equipment should be used no
closer to any part of the device, inclu-
ding cables, than the recommended
separation distance calculated from the
equation applicable to the frequency of
the transmitter.
Recommended separation distance
d = 1.2√P for 150 kHz to 80MHz
where P is the maximum output power
rating of the transmitter in watts (W) ac-
cording to the transmitter manufacturer
and d is the recommended separation
distance in meters (m).
Rated power
frequency mag-
netic fi elds
(IEC 61000-4-8)
30 A/m, 50 Hz
Power frequency magnetic fi elds should
be at levels characteristic of a typical
location in a typical commercial or
hospital environment.