- 51 -
Immunity test levels
Test
Compliance level
Electromagnetic environment - guidance
Electrostatic
Discharge
(IEC 61000-
4-2)
Contact Discharge:
±8 kV
Air Discharge: ±2 kV,
±4 kV, ±8 kV, ±15 kV
Floors should be wood, concrete or
ceramic tile. If fl oors are covered with
synthetic material, the relative humidity
should be at least 30%.
Radiated RF
EM fi led
(IEC 61000-
4-3)
80-2700 MHz; 1kHz
AM 80 %; 10 V/m
Portable and mobile RF communications
equipment should be used no closer to
any part of the device, including cables,
than the recommended separation
distance calculated from the equati-
on applicable to the frequency of the
transmitter.
Recommended separation distance
d = 1.2√P for 80 MHz to 800 MHz
d = 2.3√P for 800 MHz to 2,7 GHz
where P is the maximum output power
rating of the transmitter in watts (W) ac-
cording to the transmitter manufacturer
and d is the recommended separation
distance in meters (m).