- 52 -
Proximity
fi elds form RF
wireless com-
munications
equipment
(IEC 61000-
4-3)
385 MHz; Pulse
Modulation: 18 Hz;
27 V/m
450 MHz, FM + 5
Hz deviation: 1 kHz
sine; 28 V/m
710, 745, 780 MHz;
Pulse Modulation:
217 Hz; 9 V/m
810, 870, 930 MHz;
Pulse Modulation:
18 Hz; 28 V/m
1720, 1845, 1970
MHz; Pulse Modula-
tion: 217 Hz; 28 V/m
2450 MHz; Pulse
Modulation: 217 Hz;
28 V/m;
5240, 5500, 5785
MHz; Pulse Modula-
tion: 217 Hz; 9 V/m
Portable and mobile RF communications
equipment should be used no closer to
any part of the device, including cables,
than the recommended separation
distance 30 cm.