Maintenance...Adjustments
5-123
LTO
5.10.5 M
ANUAL
ADJUSTMENT
OF
AXIS
SETTING
5.10.5.1 P
ROCEDURE
Note: We recommend using a flat plastic lens to perform the calibration of your machine
(do not use an opthalmologic lens since it could lead to an imprecise or mistaken
calibration).
Note: You can use marked lenses (P/N 1404274) or flat plastic lenses (P/N 1000205) marked
with a line in the middle to indicate the shaft.
B
EFORE
STARTING
A
CALIBRATION
,
MAKE
SURE
THAT
THE
BLOCKING
SYSTEM
SHAFT
HAS
PREVIOUSLY
BEEN
CALIBRATED
.
Note: Block the lens on the opposite side of the marking. It will then be easier to check the
shaft error when you put the left and right lenses back to back during the final
step.
Choice of
side
Tightening of
the lens holder
shafts
Save
Error value