Symptom
Possible Causes
Recomended Solutions
Scatter is too
dim or bright or
changes over
time
Instrument is exper-
iencing large tem-
perature variation
Allow the instrument to warm up by running for 15 minutes.
Direct a fan toward the back of the instrument to dissipate excess heat, or
move the system to a temperature controlled environment.
Laser power set too high
or low
Increase or decrease the 785 SSC laser power.
Core stream position is
grossly off-center within
the flow cell due to air or
clog in the fluidics
Run the purge bubbles script from the instrument drop-down menu.
more information, see Unstable fluidics (Air or clog in system)
Large variation
in brightfield
intensity levels
Large flow speed vari-
ation due to air
Run the purge bubbles script from the instrument drop-down menu. See
solutions for unstable fluidics.
Light source delivering
variable output
Power down and power up the instrument, if this does not fix the problem,
call Luminex Technical Support.
Brightfield intens-
ity level sets
incorrectly
Intensity set before
desired flow speed has
been achieved
Allow the system to stabilize after loading a sample, and then click Set
Intensity.
Movable optics are out
of position
Power down and power up the instrument, if this does not fix the problem,
call Luminex Technical Support.
For Research Use Only. Not for use in diagnostic procedures.
51
Amnis
®
FlowSight
®
Imaging Flow Cytometer User Manual