50H / 100H Laser Systems
Message Appears: Rate high
UM-20006520DE, Rev. A
Page 56
Message Appears: Rate high
The pulse rate delivered is at least 20% more than the selected level.
• Press the
Standby
button to clear the message, then press the
Ready
button and resume normal operation. If the condition continues, turn
off the laser system for five seconds, then turn it back on. If the
condition persists, contact your Lumenis Customer Service.
Message Appears: Rate low
The pulse rate delivered is less than 80% of the selected level.
• Press the
Ready
button to clear the message. If the condition
continues, turn off the laser system for five seconds, then turn it
back on. If the condition persists, contact Lumenis Customer