Lumenis
®
Pulse
TM
50H / 100H Laser Systems
Energy Detectors Calibration
UM-20006520DE, Rev. A
Page 64
WARNING:
All personnel in the immediate area must wear eye protection rated
specifically for the Holmium laser system.
NOTE:
Optical components must be clean before the energy detectors check is
performed.
1.
Verify that all personnel are wearing appropriate laser safety eyewear.
2.
Position a calibrated, external power meter 15 cm (6 inches) from the
output end of the optical fiber.
3.
Turn on the laser system as instructed earlier in this manual.
4.
Set the laser system to deliver
5
Watts of laser energy.
5.
Target the aiming beam at the detector disc of the external power
meter.
6.
Set the laser system to
Ready
mode.
7.
Press the footswitch to deliver the laser energy into the detector disc of
the external power meter. Maintain delivery of the laser energy for
20 seconds.
8.
Release the footswitch and record the external power meter's reading.
9.
If the external power meter reading falls above or below ±20% of the
requested energy on your laser system, discontinue this procedure and
contact Lumenis Customer Service.