5.  Product Specifications 

The PEKA light engine must be operated and stored within the environmental conditions specified.

Performance specifications for individual light engines are listed on the certificate of conformance 
included with the shipping documents e-mailed to the customer.  It is important to retain the certifi-
cate of conformance for reference.  In the event that the light engine is sold, the certificate of con-
formance should be transferred to the new owner. Certificates of conformance are also recorded in 
Lumencor’s database and copies can be requested by e-mail to [email protected].  The 
request message must include the 4- or 5-digit serial number of the light engine.

Specification

Detail

Temperature

Operating

32 to 95° F (0 to 35° C)

Non-operating

-4 to 158° F (-20 to 70° C)

Humidity

Operating and non-operating

0 to 80% relative humidity, non-condensing

Altitude

Operating

0 to 10,000 feet (3,048 meters)

Non-operating

0 to 45,000 feet (13,176 meters)

Dimensions (WxLxH)

4.25 in x 6.0 in x 4.25 in  /  10.5 cm x 15 cm x 10.5 cm

Weight

3.3 lb  /  1.5 kg

Input Power Requirements

9 VDC, 4.45 A, 40 W maximum, power supply included

Power consumption

16 W at 100% intensity

Switching Speed

On/Off rate up to 1 kHz

Warm-up Period

1 s

Protection

IP Rating of X0

Sound Level

Sound Level at 1 m = 0 db(A)

Connections

BNC connection for TTL gating, DIN connector for remote control from 
Nikon Hub

Warranty

24 months parts and labor for end users

PEKA Light Engine Manual 57-10004

 

        

 

 

          

11

Summary of Contents for 90-10054

Page 1: ...Lumencor Light Engine Manual PEKA Light Engine PEKA Light Engine Manual 57 10004 lumencor com...

Page 2: ...n EN 61010 1 2010 CE Marking Low Voltage Directive 2014 35 EU EMC Directive 2014 30 EU RoHS Directive 2011 65 EU 2015 863 EU REACH Regulation EC No 1907 2006 EC EU Declarations of Conformity can be fo...

Page 3: ...uction 2 Precautions and Warnings 3 Installation and Operating Instructions 4 Spectral Output 5 Product Specifications 6 Routine Maintenance and Troubleshooting 7 Customer Support 8 Warranty PEKA Ligh...

Page 4: ...t engine is com pletely quiet and vibration free The PEKA light engine is manually controlled via knobs located on the side panel of the light engine or on the base of Nikon microscope stands Addition...

Page 5: ...corporelles Caution Statements identify conditions or practices that could result in damage to your equipment Attention d clarations qui identifient des situations ou des pratiques susceptibles d endo...

Page 6: ...les tissus le papier et les liquides animaux ou v g taux RISK GROUP 3 Warning Possibly hazardous optical radiation emitted from this product Do not look at operating lamp Eye injury may result Warnin...

Page 7: ...s optiques en verre susceptibles d tre endommag s ou d salign s par le choc r sultant d une chute sur une sur face dure DISCLAIMER Lumencor shall not be liable for injury to the user or damage to the...

Page 8: ...mencor Light Engine shall be fully evaluated to verify all applicable safety and regulatory compliance requirements prior to use When setting up the PEKA light engine be sure that the adapter Figure 1...

Page 9: ...control knob has four positions that set the light engine to deliver approximately 10 25 50 and 100 of maximum output There is no warm up time the light engine output stabilizes less than 1 second aft...

Page 10: ...lements software For control with NIS Elements the DIA LAMP must be en abled in the configuration menu 4 Spectral Output The typical spectral output distribu tion of a PEKA light engine is shown below...

Page 11: ...clude the 4 or 5 digit serial number of the light engine Specification Detail Temperature Operating 32 to 95 F 0 to 35 C Non operating 4 to 158 F 20 to 70 C Humidity Operating and non operating 0 to 8...

Page 12: ...es its return to Lumencor s facility in Beaverton Oregon for evaluation and if necessary repair Any light engine return to Lumencor for service or repair requires a material authorization RMA number T...

Reviews: