INFRATHERM pyrometer ISR 50-LO
52
Operation mode:
AAkaX
X = 1 MONO; X = 2 RATIO
Emissivity slope:
AAevXXXX
XXXX = (0800 ... 1200, decimal, equivalent to 0.800 ... 1.200)
Emissivity:
AAemXXXX
XXXX = (0010 ... 1000‰) (decimal)
Exposure time t
90
:
AAezX
X = 0 ... 6 (decimal)
0 = intrinsic time constant of the device
1 = 0.01 s
3 = 0.25 s
5 = 3.00 s
2 = 0.05 s
4 = 1.00 s
6 = 10.00 s
Clear time maximum
value storage:
AAlzX
X = 0 ... 8 (dec.)
0 = Maximum value storage off
1 = 0.01 s
4 = 1.00 s
7 = external deletion
2 = 0.05 s
5 = 5.00 s
8 = automatically deletion
3 = 0.25 s
6 = 25.00 s
9 = hold
External clearing:
AAlx
Simulation of an external deletion contact
Analog output:
AAasX
X = 0...1
0 = 0...20 mA 1 = 4...20 mA
Reading basic tem-
perature range:
AAmb
Output: XXXXYYYY (hex 8-digit, °C)
XXXX = beginning of temperature range
YYYY = end of temperature range
Reading tempera-
ture sub range:
AAme
Output: XXXXYYYY (hex 8-digit, °C)
XXXX = beginning of temperature range
YYYY = end of temperature range
Setting of tempera-
ture sub range:
AAm1XXXXYYYY XXXX (hex 4-digit) beginning of temp. range (°C)
YYYY (hex 4-digit) end of temp. range (°C)
Reading switch-of
limit:
AAaw or AAar
Output: XX (decimal) 02 … 50 in %
Switch-of limit:
AAawXX
XX (decimal) 02 … 50 in %
Reading warning
level dirty window:
AAdw
Output: XX (decimal) 00 … 99 in %
Setting warning
level dirty window:
AAdwXX
XX (dec.) ** … 99 in %, 00 … < ** switches off this function
( ** is the set value for the switch-off limit „aw“)
Address:
AAgaXX
XX = (00 ... 97)
00 ... 97 = regular device addresses
99 = global address with response
98 = global address without response (only setting com-
mands!)
Baud rate:
AAbrX
X = 0 ... 6 or 8 (dec.)
0 = 1200 baud
3 = 9600 baud
6 = 57600 baud
1 = 2400 baud
4 = 19200 baud
(7 is not allowed)
2 = 4800 baud
5 = 38400 baud
8 = 115200 baud
Changing °C / °F
AAfhX
Output: X = 0: display in °C; X = 1: display in °F
Wait time:
AAtwXX
XX = 00 ... 99 (decimal)
Internal
temperature:
AAgt
Output: XX (dec. 00 ... 98, in °C)
XXX (dec. 032 ... 208°F)
Max. internal
temperature:
AAtm
Output: XX (dec. 00 ... 98, in °C)
XXX (dec. 032 ... 208°F)
Reading signal
strength
AAtr
Output: XXXX, XXXX = 0000 ... 1500 in ‰
(calculated signal strength of the measuring object)
Error status:
AAfs
Output 1 byte hex (00 = no error)
Bit 0 = 1: Measurement unit doesn’t work
Bit 1 = 1: Internal temperature measurement doesn’t work
Laser targeting light: AAlaX
X = 0 switch off laser; X = 1 switch on laser
Reading interface:
AAin
Output: 1 or 2 (1 = RS232, 2 = RS485)
Lock keyboard:
AAlkX
X = 0 ... 3
1 = lock lk1, removal with command lk0 or power off-on
0 = removal of lock lk1
3 = continuous lock lk3, removal only with command lk2
2 = removal of lock lk3
Reading parame-
ters:
AApa
Output decimal 11-digit:
Digit 1 and 2 (10...99 or 00): Emissivity
ε
Digit 3 (0 ... 6): Exposure time t
90
Digit 4 (0 ... 8): Clear time max. storage t
clear
Digit 5 (0 ... 1): Analog output
Digit 6 and 7: (00 ... 98): Internal temperature (°C)
Digit 8 and 9 (00 ... 97): Address