INFRATHERM pyrometer ISR 50-LO 

 

 40 

7.2 

Thermal alignment  

When measuring a hot object in font of a colder background, it is usually sufficient to align the optical head 
“thermally” to achieve the highest temperature (output) signal. For this kind of alignment the pyrometer must 
be switched into 1-color mode, because in ratio mode the maximum signal can already be achieved even if 
the object fills the spot only partially.  
 
 
7.3 

Avoiding measuring errors caused by incorrect installation 

To avoid measuring errors, please note the following points when mounting the pyrometer: 

 

Infrared radiation of hot machine parts can be reflected from the measuring object and will be detected 
additionally by the pyrometer. For example if the measured object has a high reflectivity, the measured 
temperature will be mainly from the reflexions of the hot machine parts, not only from the measuring ob-
ject itself. In such a case the reflected radiation has to be blocked by suitable equipment. 

 

The spectral range of the ISR 50-LO is sensitive to bulb light or daylight (not for neon lamp light or dif-
fuse daylight). The influence of such external light must be avoided for an accurate measurement. 

 
 

8

 

Instrument settings 

 
The pyrometers are equipped with a wide range of settings for optimal adaptation to the required measuring 
condition and for getting the correct measuring temperature. 
All instrument settings can be done directly at the instrument or via serial interface and software InfraWin 
(see section 10, Settings via interface and software, user of an own communication software find all inter-
face commands on section 14, Data format UPP

®

). 

 
 
8.1 

Settings at the instrument 

The LC-display as well as the push buttons  for 
displaying and setting of the parameters are 
found inside the converter. The pyrometer is 
opened by 4 allen screws. 
 

 

 

Note:  

Only operate the device for a prolonged period of time when it is closed! After setting 
the parameters, replace the housing cover immediately to prevent dirt from entering 
the device and to meet the electromagnetic requirements (EMV). 

 

 1   PAR: With 

the 

PAR button all available parame-

ters are displayed in the following descrip-
tion (section 9). Pushing the button again 
changes the display to the next parameter 
and on the display a corresponding short 
form is displayed. 

 2  

 

 

With the arrow keys all parameter settings 
can be displayed. They are active after 
pushing the PAR button. Pushing the button 
longer changes the settings in fast mode. 

 

Push button  
short description

 

Temperature 
display or para-
meter short form

 

1

 

3

2

 

Push buttons

 3   ESC / ENT: Pushing the ESC button changes the pyrometer to measuring mode. If a parameter is 

changed with the arrow keys the indication of the ESC button changes to ENT. Pressing 
the button again confirms the value into the pyrometer. Changing the parameters again by 
pushing the PAR button doesn’t confirm this value in the pyrometer. If no button is pressed 
for 30 s the pyrometer changes to  the temperature indication without accepting the 
changed value. 

 

 

Note:  

If the pyrometer is connected to a PC via interface and the software InfraWin is 
opened with the setting window „Pyrometer Parameters“, changing of settings on the 
pyrometer is blocked. 

 
 

 

Interface switch

Push buttons

Test current key

Display

Summary of Contents for Impac INFRATHERM ISR 50-LO

Page 1: ...1 ISR 50 LO INFRATHERM Pyrometer Betriebsanleitung User Manual IMPAC Spezialist f r ber hrungslose Temperaturmessung IMPAC Specialist in non contact thermometry www imaxsun com...

Page 2: ...5 2 1 Short description 36 5 3 Connecting the pyrometer to a PC 36 5 4 Connecting to RS232 interface 37 5 5 Connecting to RS485 interface 37 5 6 Connection of additional analyzing devices 37 6 Mechan...

Page 3: ...mum internal temperature MaxIntTemp 45 10 Settings via interface and software 45 10 1 Installation 45 10 2 Program start 45 10 3 The menu 46 10 4 Beginning 46 10 5 Number of devices 46 10 6 Basic sett...

Page 4: ...arameter settings Parameters Adjustable or readable at the instrument or via interface Measuring temperature operation mode ratio mono emissivity slope K or emissivity exposure time t90 clear times tc...

Page 5: ...late the temperature This technique offers the following advantages compared with the standard one color pyrometers The temperature measurement is independent of the emissivity of the object in wide r...

Page 6: ...Only qualified specialists are allowed to connect such devices to the mains voltage 5 Electrical Installation The ISR 50 LO is powered by a voltage of 24 V DC possible range 18 36 V DC With the conne...

Page 7: ...g is frozen as long as pin 7 24 V is connected to pin 5 Pin 8 6 If the pyrometer is used in ratio mode a relay contact for the integrated contamination alarm is available It switches at the too low si...

Page 8: ...rate has to be reduced by 50 when the transmission distance is doubled see also 9 9 Baud rate 5 6 Connection of additional analyzing devices Additional analyzing instruments for example a LED digital...

Page 9: ...radius 0 6 mm fibre 0 2 mm fibre 0 1 mm fibre green mark red mark yellow mark for short time max 250 C 30 mm 50 mm 50 mm permanent max 250 C 50 mm 120 mm 120 mm wound up Zustand max 50 C 50 mm 120 mm...

Page 10: ...ce possible range is 340 4500 mm the measuring distance is always measured from the front of the lens Only in this distance the spot sizes mentioned in the table 7 1 will be achieved A tape can be use...

Page 11: ...e user of an own communication software find all inter face commands on section 14 Data format UPP 8 1 Settings at the instrument The LC display as well as the push buttons for displaying and setting...

Page 12: ...Adress Adr 00 Baud rate kBaud 19 2 kBd Wait time tw for RS485 10 Temperature display C F C Switch for interface RS485 RS232 RS232 9 Parameters settings 9 1 Operating mode Mode The standard use for the...

Page 13: ...same temperature Different materials have different emissivities ranging between 0 and 100 settings at the pyrometer between 5 and 100 Additionally the emissivity is depending on the surface condition...

Page 14: ...Here the maximum value for each object has to be indicated In this mode the maximum value is stored until a new hot object appears in the measuring beam The temperature which has to be recognized as h...

Page 15: ...nal in such a high level that a correct temperature measurement will be impossible this can also happen if the object is smaller than the spot size of the pyrometer To avoid wrong measurements in adva...

Page 16: ...he operating and analyzing software InfraWin is included in delivery of the pyrometer With this software all pyrometer functions also can be used on the PC except changing the interface or using the t...

Page 17: ...sing RS485 may be operated simultaneously by the same interface if two different addresses have been properly entered see 9 8 Address 10 5 Number of devices With a click on Number of devices InfraWin...

Page 18: ...ngth of the relative signal warning level graphically and numerically switch off level graphically and numerically The measuring mode the emissivity or emissivity slope and values for the warning leve...

Page 19: ...nce the temperature entry has been entered and confirmed with OK InfraWin will then calculate the emissivity slope which occurs with the new temperature This is displayed immediately and can be used f...

Page 20: ...e mouse Selecting the Trend output initially causes all the saved data to be displayed If the data exceeds an amount that can be represented reasonably you may Zoom in on a partial segment using the m...

Page 21: ...ter is designed for a storage temperature of 20 60 C with non condensing conditions Other kind of storage can damage or malfunction the pyrometer 12 Maintenance 12 1 Safety Attention during pyrometer...

Page 22: ...are see 10 6 Basic settings Test commands can be exchanged directly with the pyrometer The data exchange occurs in ASCII format with the following transmission parameters The data format is 8 data bit...

Page 23: ...AdwXX XX dec 99 in 00 switches off this function is the set value for the switch off limit aw Address AAgaXX XX 00 97 00 97 regular device addresses 99 global address with response 98 global address w...

Page 24: ...ough for this 2 The pyrometer s response will follow after 3 ms at latest 3 If there is no response there is a parity or syntax error and the inquiry has to be repeated 15 Reference numbers 15 1 Refer...

Reviews: