IS 12-TSP / IGA 12-TSP Manual 

Settings / parameter descriptions 

  24 

4.6

 

Clear time of the maximum value storage (t

CLEAR

If the maximum value storage is switched on always the highest last 
temperature value will be displayed and stored. The storage has to be cleared 

at regular intervals for exchanging by a new and actual value.  

This feature is particularly useful when fluctuating object temperatures cause 

the display or the analog outputs to change too rapidly, or the pyrometer is 
not constantly viewing an object to be measured. In addition, it may also be 
beneficial to periodically delete and reset the stored maximum values. 

The following settings are possible: 

 

OFF: 

At clear time “OFF“ the max. value storage is switched off and only momentary 
values are measured. 

0.01...25 s:

 If any clear time between 0.01 s and 25 s is set, the maximum value is estimated and 

held in double storage mode after the entered time, the storage will be deleted. 

extern:

 

The external clearing can be activated and used within an own software (see section 

7, Data format UPP

) or via an external contact (for connection see 

3.1.1 

Connector pin assignment

 on the back side of the pyrometer). In this case, the 

storage operates only in single storage, because only a single deletion mechanism is 
used. 

  

auto:

 

The “auto” mode is used for discontinuous measuring tasks. For example, objects 
are transported on a conveyer belt and pass the measuring beam of the pyrometer 

only for a few seconds. Here the maximum value for each object has to be indicated. 
In this mode the maximum value is stored until a new hot object appears in the 

measuring beam. The temperature which has to be recognized as “hot“ is defined 
by the low limit of the adjusted sub range. The stored maximum value will be 

deleted when the temperature of the new hot object exceeds the low limit “from“ 
of the sub range by 1% or at least 2 °C. If a lower limit is not entered, the maximum 

value storage will be deleted whenever the lower level of the full measuring range 
has been exceeded.

  

hold:

 

The function “hold” enables to freeze the current temperature reading at any 

moment. For this an external push button or switch has to be connected, which 
holds the temperature reading as long as the contacts are closed. 

Operation Note:

 

Dependent on the settings the maximum value storage either works in 
single storage mode or in double storage mode:

 

Single Storage 
Mode:

 

The single storage is used when you want to reset the stored value using an 
external impulse via one contact closure from an external relay (i.e. between 

two measured objects). The relay contact is connected directly to the 
pyrometer between pins J and K. This mode allows a new value to be 

established, after each impulse from the reset signal

.   

 

 

 

 

Settings: 

0.00 s (= OFF) 

0.01 s 

. . . 

25 s 

extern 

auto 

Hold 

Summary of Contents for IGA 12-TSP

Page 1: ...MANUAL IMPAC Pyrometer IS 12 TSP IGA 12 TSP...

Page 2: ...espective holders Service Centers LumaSense Technologies Inc North America Sales Service Santa Clara CA USA Ph 1 800 631 0176 Ph 1 408 727 1600 Fax 1 408 727 1677 LumaSense Technologies GmbH Other Tha...

Page 3: ...for the connector on the back of the pyrometer 13 3 2 Connecting the pyrometer to a PC 14 3 2 1 Connecting to RS232 interface 14 3 2 2 Connecting to RS485 interface 15 3 3 Connection of additional ana...

Page 4: ...s 25 4 12 kBaud baud rate 26 4 13 10 mA test current output 26 4 14 Wait time tw 26 4 15 Switch contact 26 5 Software InfraWin 27 5 1 Connecting the pyrometer to a PC 27 5 2 Installation 27 5 3 Progra...

Page 5: ...ment Security note laser beam Indicates to the danger of a built in laser targeting light MB Shortcut for Temperature range in German Messbereich 1 1 2 Terminology The terminology used in this manual...

Page 6: ...cts caused by LumaSense Technologies The Windows compatible software was thoroughly tested on a wide range of Windows operating systems and in several world languages Nevertheless there is always a po...

Page 7: ...0176 Email support lumasenseinc com Frankfurt Germany Telephone 49 0 69 97373 0 Email support lumasenseinc com Erstein France Telephone 33 0 3 88 98 98 01 Email support lumasenseinc com 1 6 Shipments...

Page 8: ...e in humid areas or shipment overseas the device should be placed in welded foil ideally along with silica gel to protect it from humidity The pyrometer is designed for a storage temperature of 20 to...

Page 9: ...400 C 600 3000 C 300 1700 C 850 2520 C Sub Range Any range adjustable within the temperature range minimum span 51 C Signal Processing Photoelectric current digitized Spectral Range IS 12 TSP 0 94 m a...

Page 10: ...missivity Response time t90 Clear times tCL of the maximum value storage Temperature sub range Analog output 0 to 20 or 4 to 20 mA Switch points for limit contacts C or F Interface RS232 or RS485 Addr...

Page 11: ...SP 300 1700 360 Note The calibration adjustment of the instruments was carried out in accordance with VDI VDE directive Temperature measurement in industry Radiation thermometry Calibration of radiati...

Page 12: ...as temperature or parameters 3 Parameter indicator 4 Parallax free thru lens view finder 5 Setting keys operate with tipp of ball point pen 6 Connection cable 7 Laser warning label only instruments wi...

Page 13: ...imum value storage via contact see chapter 8 Reference numbers and 12 pin angle connector The cable includes a short RS232 adapter cable with a 9 pin SUB D connector for direct PC communication This a...

Page 14: ...able at the pyrometer for connection to a processor PC Standard on a PC is the RS232 interface At this interface one pyrometer can be connected if the interface is set to RS232 Only short distances ca...

Page 15: ...nection of additional analyzing devices Additional analyzing instruments for example a LED digital display instrument only needs to be connected to a power supply and the analog outputs from the pyrom...

Page 16: ...out temperature measurements The following points must also be observed in the special case of the TSP device 3 4 2 Special aligning notes The size of the spot size is determined first and foremost by...

Page 17: ...ex works This is the standard setting for TS pyrometers which is achieved with maximum precision at minimum noise values Changing this setting is only necessary if the pyrometer is to be used for othe...

Page 18: ...17 The aperture is the effective lens diameter It is depending on the objective length The biggest aperture value 17 mm belongs to the fully extended objective S 30 mm the smallest aperture value 13...

Page 19: ...for each of the 3 focusable optics This corresponds to the longest or the shortest objective length As an example a further value between max and min is shown The objective length S can be measured wi...

Page 20: ...IS 12 TSP IGA 12 TSP Manual Controls and Installation 20 To ensure consistent document formatting this page was intentionally left blank...

Page 21: ...2 kBd Interface RS RS232 Limit contacts end of temperature range Hysteresis 2 C Test current output 10 mA off 4 2 Instrument settings All instrument settings can be done directly at the instrument To...

Page 22: ...lay to the next parameter after the last parameter it changes to the actual temperature reading With the arrow keys and all parameter settings can be displayed Pushing the button longer changes the se...

Page 23: ...1 1 Steel heavily scaled 0 92 0 93 0 85 to 0 9 Steel rolling skin 0 89 0 88 0 8 to 0 88 Steel molten 0 35 0 3 0 2 to 0 25 Slag 0 85 0 85 0 8 to 0 85 Aluminum bright 0 13 0 15 0 1 Chromium bright 0 32...

Page 24: ...is used for discontinuous measuring tasks For example objects are transported on a conveyer belt and pass the measuring beam of the pyrometer only for a few seconds Here the maximum value for each obj...

Page 25: ...cted according to the signal input of the connected instrument controller PLC etc Note Settings for Analog Output 0 to 20 mA or 4 to 20 mA 0 to 20 mA 21 Low or 4 to 20 mA 21 Low setting 21 Low Analog...

Page 26: ...lly after 1 minute and the instrument is working in the measuring mode 4 14 Wait time tw Using a pyrometer with RS485 it is possible that the connection is not fast enough to receive the pyrometer s a...

Page 27: ...nc com 5 1 Connecting the pyrometer to a PC The program InfraWin can operate up to two devices Two devices using RS485 may be operated simultaneously by the same interface if two different addresses h...

Page 28: ...IS 12 TSP IGA 12 TSP Manual Software InfraWin 28 To ensure consistent document formatting this page was intentionally left blank...

Page 29: ...ptics These optics can be changed against another e g if the lens is scratched or the pyrometer will be used for other measuring distances When changing the optics a recalibration of the pyrometer is...

Page 30: ...IS 12 TSP IGA 12 TSP Manual Maintenance 30 To ensure consistent document formatting this page was intentionally left blank...

Page 31: ...ive settings only at setting commands not at enquiry commands Example Entry 00em CR The emissivity setting of the device with the address 00 is returned Answer 0970 CR means Emissivity 0 97 or 97 0 De...

Page 32: ...00 Baud 4 19200 Baud 7 is not allowed 2 4800 Baud 5 38400 Baud 8 115200 Baud 3 9600 Baud 6 57600 Baud Wait time AAtwXX XX 00 99 decimal Error status AAfs Output XX XX 00 FF 00 no error 01 FF error cod...

Page 33: ...ce type software version AAve Output XXYYZZ 6 digit decimal XX 07 instruments ID for IS 12 TSP or IGA 12 TSP YY Month of software version ZZ Year of software version Software version in detail AAvs tt...

Page 34: ...IS 12 TSP IGA 12 TSP Manual Data format UPP Universal Pyrometer Protocol 34 To ensure consistent document formatting this page was intentionally left blank...

Page 35: ...ngth 25 m 90 connector 3 820 550 Connection cable length 30 m 90 connector 3 820 740 Connection cable length 5 m 90 connector temperature resistant up to 200 C 3 821 120 Additional cable for limit con...

Page 36: ...IS 12 TSP IGA 12 TSP Manual Reference Numbers 36 To ensure consistent document formatting this page was intentionally left blank...

Page 37: ...place it see 6 2 6 3 Temperature indication too high Emissivity set too low Set lower correct emissivity corresponding to the material see 4 4 The measurement is influenced by reflections of hot machi...

Page 38: ...IS 12 TSP IGA 12 TSP Manual Troubleshooting 38 To ensure consistent document formatting this page was intentionally left blank...

Page 39: ...al 13 Interface 25 L Language 27 Laser targeting light 15 22 Legend 5 Liability 6 M Maintenance 29 P Parameter display 22 Parameter indicator 22 Physical User Interface 12 R Reference numbers Accessor...