IS 12-TSP / IGA 12-TSP Manual
Introduction
•
10
Environment
Protection Class:
IP65 (DIN 40 050)
Ambient Temperature:
0 to 60 °C at housing
Storage Temperature:
-20 to 70 °C
Relative Humidity:
Non condensating conditions
Weight:
2.2 kg
CE-Label:
According to EU directives about electromagnetic immunity
Interface
Display:
Built-in 5 digit LED display, height 13 mm, additional function
LED’s
Control Panel:
4 keys, operate with tip of ball-point pen
Sighting:
Built-in parallax free thru-lens view finder; additionally, laser
targeting light (max. power level < 1 mW,
λ
= 630 - 680 nm,
CDRH class II)
(at IS 12-TSP with temperature range 850 - 2520 °C only with
view finder)
Parameters:
Adjustable at the instrument or via serial interface:
Emissivity
ε
; Response time t
90
; Clear times t
CL
of the maximum
value storage; Temperature sub range; Analog output 0 to 20
or 4 to 20 mA; Switch points for limit contacts; °C or °F;
Interface RS232 or RS485; Address; Baud rate; Test current
output
Additionally (only via interface adjustable):
Key lock, wait time, Recalibration (with
special software)
Communication
Analog Output:
0 to 20 mA or 4 to 20 mA switchable
Test Current Output:
Fixed 10 mA
Digital Interface:
Switchable at the instrument: RS232 or RS485 addressable (half
duplex), baud rate 2.4 up to 115.2 kBd
Maximum Value Storage:
Built-in single or double storage. Clearing with adjusted time t
clear
(off; 0.01 s; 0.05 s; 0.25 s; 1 s; 5 s; 25 s), extern, via interface or
automatically with the next measuring object
Limit switches:
2 relay outputs (switch-over relay contacts), switch power
max. 30 W (I
max
: 1 A, U
max
: 60 V DC)
Rise and fall times: exposure time of the pyr 2 ms