IS 12-TSP / IGA 12-TSP Manual
Settings / parameter descriptions
•
21
4
Settings / parameter descriptions
All setting keys and switches for operation the instrument are placed inside the
housing and can only be used after removing the cover (four screws). Ex works the
instrument is operating in pouring stream mode (setting of other modes only via
software InfraWin, see Chapter
5
,
Software
InfraWin
).
4.1
Factory Settings
Temperature display (°C / °F) = °C
Emissivity (
ε
) = 1.000
Exposure time (t
90
/ s) = 1 s
Clear time (t
Clear
) = 0.00
Analog output (
0 / 4 mA
) = 0 ... 20 mA
Sub range (
from / to
) same as temperature range
Address (
ADR
) = 00
Baud rate (
kBaud
) = 19.2 kBd
Interface (
RS
) = RS232
Limit contacts = end of temperature range
Hysteresis = ±2°C
Test current output (
10 mA
) = off
4.2
Instrument settings
All instrument settings can be done directly at the instrument. To activate the adjusting keys, a
tip of ball pen must be used. This avoids changing parameters by mistake.
Temperature-
or
parameter
display
Targeting light on / off
Setting keys:
1) Parameter selection
2) Change of parameters:
3) Confirmation of parameter
Indication of limit switches
Indication °C or °F
Parameter indicator
Emissivity
ε
Exposure time (t
90
/s)
Clear time of max. value storage (t
CL
/s)
Beginning of sub range (FROM)
End of sub range (TO)
Analog output 0 or 4 ... 20 mA (0/4 mA)
Setting of limit switches (SETP)
Temperature in °C or °F (°C/°F)
Interface RS232 or RS 485 (RS)
Address (ADR)
Baud rate (kBaud)
Test current output (10 mA)