Operating Manual CHM 15k
R13 / 07-2019
Cleaning, Maintenance and Service Instructions
66
Interval
Preventive maintenance measures
Comment
Regular checks
Check the dehumidifier bag
CONTAINER DRI II for integrity and replace if
necessary
By service staff only
At least once a year
Change the dehumidifier bag CONTAINER
DRI II
By service staff only
3 to 5 years
Replace LOM, if signs of laser wear, (falling
repetition rate), replace SD card
By service staff only
Approx. every 5 years
As a preventive measure:
Replace the rubber seal of the inner door (in
case of material fatigue)
By service staff only
Approx. every 5 years
As a preventive measure:
Replace the lightning conductor (also after
lightning strike)
By service staff only
Approx. every 8 years
As a preventive measure:
Replace the electronic mainboard
By service staff only
Table 22 Preventive maintenance intervals and measures.