Operating Manual CHM 15k
R13 / 07-2019
Communication via RS485 & Ethernet
46
Group
Hex code
Error description
Duration [s]
xxxx xxxC
Mainboard identifier cannot be read
∞
xxxx xxxD
Firmware does not match the CPU version
∞
2
Data transmission and storage
xxxx xx0x
Data transmission and storage working properly
xxxx xx1x
Faulty file system on SD card repaired
60
xxxx xx2x
RS485 baud rate / transfer mode were reset
60
xxxx xx3x
AFD problem
60/ 600
xxxx xx4x
RS485 data telegram cannot be transmitted
16
xxxx xx5x
RS485 data telegram cannot be generated
16
xxxx xx6x
Write error in NetCDF file
60
xxxx xx7x
New NetCDF file cannot be generated
60
xxxx xx8x
SD card not available or defective
∞
3
Temperatures
xxxx x0xx
Temperature values are in range
xxxx x1xx
Detector temperature not in best performance range
(setpoint -1 °C ... +3 °C)
60
xxxx x3xx
Laser optical module (LOM) temperature outside the
valid range (25 °C ... 49 °C)
60
xxxx x4xx
Internal temperature outside the working range
(5 °C ... 50 °C)
16
xxxx x5xx
External temperature outside the working range
(-35 °C ... 50 °C)
60
xxxx x6xx
Laser optical unit temperature control disabled due to
safety reasons
16 /
∞
xxxx x7xx
Laser controller temperature too high
60
xxxx x8xx
Laser head temperature too high or too cold
16
xxxx x9xx
Laser optical unit temperature too high
16
xxxx xAxx
Laser temperature out of working range or invalid
delete
*
4
Calculation / processing in sky condition algorithm
xxxx 0xxx
Data processing OK
xxxx 1xxx
Problem in calculating the visual range
16 / 60
xxxx 2xxx
Problem in calculating of aerosol layers
60
xxxx 3xxx
Problem in calculating of cloud coverage
60
xxxx 4xxx
Problem in calculating of clouds
60
xxxx 5xxx
Abnormal signal
60
xxxx 6xxx
Incorrect dimensioning of raw data
16
xxxx 7xxx
No new data
16