PyroTec™ PRO Hardw are User’s Manual, V1.2
93
RT-MN024-1.2 02/21
WARNING
Risk of fire or explosion
If inflammable reagents were used in the process, remains of these substances on the
waste DiTis may accumulate and form combustible vapors.
•
If inflammable reagents are in use, change the DiTi waste bag frequently.
•
Perform a risk assessment to define further measures.
The filling height of the DiTi waste bag must be checked regularly. Make sure that there is
no DiTi jam within the DiTi waste slide and change the DiTi waste bag at least once at the
end of the day.
Removal
Follow the procedure below to change the DiTi waste bag.
Changing the DiTi waste b ag
•
Lift the fastener to remove the bag housing.
Note
: Be sure to dispose of waste according to your laboratory guidelines.
•
Remove the DiTi waste bag and dispose of it appropriately.
Installation
•
Install a new DiTi waste bag into the empty bag housing.
Note
: The waste bag must be suitable of disposable tips, and in case you are oper-
ating with biohazardous material suitable for that material as well, it must e.g.
have an adequate thickness and be labeled with a corresponding biohazard label.
A
DiTi waste bag
B
Bag housing
C
Fastener
D
DiTi waste slide