PyroTec™ PRO Hardw are User’s Manual, V1.2
89
RT-MN024-1.2 02/21
Replacing DiTi
Adapter
This section describes the replacement of the DiTi adapters.
Preparation
To prepare for replacement, proceed as follows:
•
Switch the instrument off.
•
Open the front safety panel.
•
Manually move all Z-rods up to their topmost position.
•
Move all Z-rods together towards the front of the instrument.
•
Spread the Z-rods all the way.
Removal
To remove the DiTi adapter, proceed as follows:
•
Hold the tip ejector tube while unscrewing the DiTi cone, using the supplied cone
wrench.
•
Remove the tip ejector tube.
•
Unscrew the adapter cylinder
•
Pull the tubing extension and the pipetting tube approximately 25 mm (1 in) out of the
tip adapter.
•
Separate the tubing extension from the pipetting tubing.
•
Remove the tubing extension together with the adapter cylinder.
Installation
To install the DiTi pickup mechanism, proceed as follows:
DiTi pick up mechanism installation
A
E
D
C
F
I
H
G
B
6-8 mm
.24-.32 in.
J