PyroTec™ PRO Hardw are User’s Manual, V1.2
26
RT-MN024-1.2 02/21
Serial Number
Label
The identification data is also printed on the serial number label (B). This label can be
viewed from the front side of the instrument and is attached below the diluters.
Instrument
Heights
Instrument equipped with:
Instrument height:
Air displacement pipetting arm, Air LiHa
1070 mm
42.1 in.
Robotic manipulator arm, RoMa
920 mm
36.2 in.
Safety Panel
Opening
Dimensions
Dimensions of the opening in the front safety panel:
PyroTec™ PRO: 1130 x 170 mm (44.5 x 6.7 in.)
Supply Ratings
PyroTec™ PRO
Line voltage (single phase)
100 - 120, 220 - 240 V AC
(-15% / +10%)
Frequency
50/60 Hz
Power
1200 VA
Fuses
2 x T10A (instrument power)
•
x T2A (main powered op-
tions)
Electrical
Safety
Note:
Considering possible under voltage on the mains supply, the combined power con-
sumption in the lower input voltage band (100 - 120 VAC) must not exceed 1000 VA in or-
der to keep the input current below the fused values.
Classification with regard to electrical safety according to EN/IEC standards:
Overvoltage category
II
IEC 60664-1
Pollution degree
2
(EN) IEC 61010-1
Power Switch
The power switch is placed at the level of the front access panel. The power switch does
not switch the mains voltage directly, but gives a control signal to the power supply.
Specification
Description
Circuit break
By unplugging the instrument.
Power on delay
0.2 - 0.5 sec.
Power off delay
1 - 2 sec.