D9U001MCX-0101_03
95
14 Scales Control
Multicare X is equipped with a weighing system that allows weighing the patient in bed. The control and display elements for the
weighing system are situated on the Scales screen on the Multiboard X LCD touchscreen.
1. Weight Value Indicator – Current Weight
2. NEW PATIENT Icon (press to enter NEW PATIENT Guide)
3. Scale Interval Switch Icons (0,5kg/0,1kg)
4. HOLD Icon (press to enter HOLD Mode Guide)
5. Stabilized Scales Indicator
6. Scales Screen Icon
7. ZERO/T Icon (press to enter Guide for Taring)
8. Memory Icon (save weight value)
Fig.
Scales Screen - LCD touchscreen
25. 12. 25
22:58
CURRENT WEIGHT
Max 250 kg, Min 10 kg, e = 0,5 kg
0,5
0,1
115
,5 kg
ZERO
T
09
1
2
6
8
7
4
5
3
14.1 Preparation
CAUTION!
Incorrect use of scales due to incomplete preparation!
► Before each patient admission tare the scales.
► Install mattress and accessories to prepare bed before patient admission and using the scales.
14.2 Taring
Taring can be done in a range of 5kg to 249.5kg. Taring is used to set “0” on the display before placing the patient on the bed.
Taring must be done with an unloaded bed with mattress, bed sheets, pillows and necessary accessories, without the patient. It is
recommended to position the mattress support platform about 20 cm above the lowest horizontal position.