D9U001MCX-0101_03
56
Status (Pop-up window)
Meaning
How to change the status
Bed is overloaded by more than 250 kg.
Safe Working Load is exceeded by more
than 10 kg.
Remove load!
OptiCare X System Control Unit (SCU) is
disconnected.
Contact service department
approved by manufacturer!
Automatic CPR deflation is not available.
OptiCare X System Control Unit could be
disconnected.
Use manual CPR valve on a
mattress side to deflate the Opti
-
Care X mattress!
System fault of the Multicare X bed!
Stop using the Multicare X bed
and contact service department
approved by manufacturer!
Scales system is disconnected and its
functions are disabled! Bed Exit Alarm
Monitoring and automatic pressure
settings of the OptiCare X mattress are
disabled!
Contact service department
approved by manufacturer!
Column unit fault!
Contact service department
approved by manufacturer!
Bed overloaded.
Remove load from the bed.
>250 kg
B123
Pump is disconnected. Call service.
OK
SERVICE
B200
Mattress disconnected.
Use manual CPR.
B125
System fault.
Stop using bed. Call service.
SERVICE
B111
Scale system and BEA disconnected. Auto
pressure settings disabled. Call service.
OK
B202
SERVICE
Columns fault.
Call service.
SERVICE
B121-x