9 Power Cord
Attachment plug is means of connecting and disconnecting bed from the mains.
Mains power cord must be attached with a hook at the head end of the bed during transport.
CAUTION!
Disconnecting bed from the mains does not stop motions of the bed!
►
Stop the bed before disconnection bed from the mains.
Where the integrity of the external protective conductor in the installation or its arran-
gement is in doubt
►
operate the bed from internal accumulator only.
Fig.
Hook for hanging power cord
8.8 Firmware
The bed includes firmware that can be updated only by an authorised service technician.
This firmware is protected against unauthorised access by mechanical housing (tool is needed to access), by seal (com
-
ponents with processor are sealed), by exclusive compatibility with an authorised software tool and by check of compati-
bility of the new firmware with the bed.
Fig.
Bed Transport Mode (Attendant Control Panel)