
D9U001MCX-0101_03
73
12.2.1 Bed and Keyboards Illumination
Bed and Keyboards Illumination helps the nursing staff as well as the patient to orientate. The illumination is turned off when the
bed is powered by accumulator.
The Multicare X bed is equipped with three-phase illumination:
1. Full intensity of illumination
2. Lowered intensity of illumination
3. Illumination is turned off
The lowered intensity of illumination is set up after the bed is connected to the mains.
After pressing the GO button:
►
The Patient Control Panels and Handset keyboards and Bed Illumination will light up at full intensity.
After elapsing the GO period:
►
The illumination intensity of Patient Control Panels and Handset keyboards will be lowered.
7 minutes after GO period has elapsed:
►
The intensity of Bed Illumination will be lowered.
In the event any button is pressed on the Attendant Control Panel, Handset or Patient Control Panels outside of GO peri-
od:
►
The Bed Illumination will light up at full intensity for 10 minutes and after that the illumination will be lowered.
►
Simultaneously, the Patient Control Panels and Handset keyboards will light up at full intensity for 7 seconds. Then the
illumination will be lowered.
After activating the bed via any Foot Control:
►
The full intensity of Bed Illumination will be activated for 9 minutes and 40 seconds.
In case of pressing any Foot Control once outside the period of activation:
►
All the Patient Control Panels and Handset keyboards will light up at full intensity for 7 seconds and then the illumination
will be lowered.
►
The full intensity of Bed Illumination will be activated for 10 minutes and then the illumination will be lowered.
Turning off all illumination:
►
Press
Illumination Control Button
(
7
) on the Attendant Control Panel and all illumination on the bed (Patient Control
Panels and Handset keyboards and Bed Illumination) will be turned off. This function is not blocked by the preceding activation by
the GO button.