D9U001MCX-0101_03
51
Status (Pop-up window)
Meaning
How to change the status
Function locked.
Unlock function if needed. Enter
the Lock Screen to unlock the
selected function.
GO Button not activated. Use of the cont
-
rol elements is disabled.
Press GO Button to activate cont
-
rol elements if needed.
Lateral Tilt disabled when siderail folded
down or a column unit is disconnected.
Raise siderail up to allow Lateral
Tilting to continue.
Incorrectly inserted X-Ray Cassette Hol
-
der. Backrest Adjustment is disabled.
Insert X-Ray Cassette Holder
correctly!
Risk of collision. Positioning blocked to
avoid collision of the bed with floor or
collision of the bed with bed equipments.
To continue in positioning, adjust
bed extension to shorten mattress
support platform.
Backrest blocked.
Insert X-ray cassette holder properly.
Collision warning.
Shorten bed platform.
12.1.2 Statuses (Multiboard X - LCD touchscreen)
Positioning locked.
Unlock the function in lock screen.
Positioning blocked.
Press GO button.
Lateral tilt blocked.
Lift all siderails up.
There are 3 types of pop-up windows according to the coloured stripe in the upper part of pop-up window.
Red colour indicates warning with required action. Orange colour indicates caution with recommended action. Grey colour indicates
only notification.