Step 1: Cleaning
1. Unplug the mains (AC power source) before cleaning and disinfection.
2. As necessary, first remove visible soil from the lift with a cloth moistened with warm water and a neutral, approved
cleaner/disinfectant. See "Application of commonly used Cleaning / Disinfectants on Liko products." Do not use a cloth that is
dripping wet.
•
A soft bristle brush may be used for hard-to-clean areas to remove stains and resistant dirt and to loosen hardened soil.
•
Use as many wiping cloths as needed to remove the soil. Replace cloth when soiled.
3. Wipe down the entire lift starting from the top down. Give special attention to seams, cracks and other areas where soil
may accumulate. In particular, pay special attention to the following areas:
NOTE! Do not clean the piston rod!
• Sling bar
• Mechanical emergency lowering
• Handles
• Control box
• Battery
• Hand control
• Emergency stop
• Operation panel/display (where applicable)
• Lever for base width adjustment (where applicable)
• Pedal for base width adjustment (where applicable)
• Locking handles
• Wheels
Cleaner/Disinfection:
NOTE:
It is important to remove all visible soil from all areas before continuing to remove non-visible soil.
With a new wiping cloth soaked in an approved cleaner/disinfectant, use firm pressure to wipe all surfaces of the lift.
Use a new or clean wiping cloth as often as necessary. Make sure the following items are cleaned:
•
Hand Control
•
Sling (refer to the specific sling Instruction for use and 7EN160884 Care and Maintenance of Liko Slings
•
Lift
•
Power cord
•
Slingbar
•
Scale (if applicable)
Damaged items should be replaced!
Step 2: Disinfection:
1. For the use of suitable disinfectants see "Application of commonly used Cleaning / Disinfectants on Liko products” in this
document.
2. Follow the manufacturer's instructions.
3. Make sure all surfaces r
emain wet with the cleaner/disinfectant
for the
specified contact time
. Rewet surfaces with a new
wiping cloth as necessary and per the manufacturer's instructions.
NOTE:
If bleach is used with another cleaner/disinfectant, use a new or clean cloth/wipe soaked in tap
water to remove any disinfectant residue prior to and after the bleach application.
The lift may not be cleaned with CSI or equivalent.
The hand control may not be cleaned with Viraguard or equivalent.
The control box may not be cleaned with Anioxy Spray or equivalent.
Do not clean!
(piston rod)