Troubleshooting
The lift do not work up/down with
Hand Control.
The base-width adjustment
doesn’t work (in/out) with the
Hand Control.
The charger doesn’t work.
The lift is stuck in the high position.
If you hear unusual sound from the lift.
1. Make sure that the emergency stop button has not been activated
(shall not be pressed in).
2. Check that the charger cables are connected correctly.
3. Make sure that the battery is properly seated in the control box.
4. Try an alternative mains outlet.
5. If the problem persists, please contact Hill-Rom.
1. Make sure that the emergency stop button has not been activated
(shall not be pressed in).
2. Make sure that the battery is properly seated in the control box.
3. Check the battery capacity.
4. Check that the handcontrol cable is connected correctly.
5. Electrical emergency lowering, use the operation panel to lower
the patient onto a firm surface, see chapter; "Operation".
6. Use the mechanical emergency lowering device to lower the
patient onto a firm surface, see chapter; Operation.
7. If the problem persists, please contact Hill-Rom.
1. Make sure that the emergency stop button has not been activated
(shall not be pressed in).
2. Check the battery capacity.
3. Make sure that the battery is properly seated in the control box.
4. Check that the charger cable is not connected to an electric outlet.
5. Check that the handcontrol cable is correctly connected to the
control box.
6. Check that the lift arm actuator cable is correctly connected to the
control box.
7. Check that the base widht actuator cable is correctly connected to
the control box.
8. If the problem persists, please contact Hill-Rom.
Contact Hill-Rom.