Get Started
248083296-1 Rev 3
3-5
Inlet Nozzle
Note:
The user is
responsible for
maintaining a nominal
flow rate of 100 ml/min +
5% through the sensor.
The sample source is attached using ¼" flare tubing to the Inlet Nozzle.
The user is responsible for maintaining a nominal flow rate of 100 ml/
min + 5% through the sensor to insure the accuracy of the data.
If there are any questions or problems during system setup, please
contact Lighthouse Worldwide Solutions’ Technical Support at
[email protected].
Figure 3-4 Example of an RLPC Installation
Sample Source
Valve
Valve
Liquid Flow
Meter
Pump
(Optional)
Sample Return
or to Drain
Inlet
Nozzle
Outlet
Nozzle
RLPC
sensor