Programming
248083296-1 Rev 3
4-7
Range2= 1000.
Service High = 3000 mV
Service Low = 145 mV
Alarm Channel = 1
Alarm Threshold = 1000 counts
Suppress Alarms = 2
Channel 1 = 0.3
Channel 2 = 0.5
5. Type:
<m1>
-- the unit
’
s menu is displayed to show
commands to use to change parameters.
<m1>
*** User Menu ***
<?> Display Parameters
<V> Verbose Mode
<sa> Start Counting
<sb> Stop Counting
<gv> Get Version Number
<ra#> Set Range 1 (counts)
<rb#> Set Range 2 (counts)
<ta#> Set Sample Seconds
<aa#> Set Alarm Channel #
<ab#> Set Alarm Threshold (# of counts)
<ae#> Set Alarm Suppression #
6. The instrument will wait 20 seconds after the last command is sent,
then begin sampling
ASCII
Programming
Syntax
Using a simple ASCII protocol, the REMOTE RL32 family of
instruments can be programmed from an ASCII terminal (i.e. Hyper
Terminal). The protocol format is based on a start character, followed
by a command, which is then followed by a terminating character. Any
characters before the start character, or after the terminating character
are ignored. White spaces between command character and command
parameter are ignored.
Command Structure
The commands are defined as single case-sensitive ASCII characters.
Format is defined as:
Summary of Contents for RLPC 0.1mm
Page 1: ......
Page 2: ...00...
Page 3: ...Lighthouse Worldwide Solutions REMOTE Liquid Particle Counters 4 20mA Output Operating Manual...
Page 10: ...Lighthouse REMOTE Liquid Particle Counter 4 20mA output Operating Manual ii 248083296 1 Rev 3...
Page 20: ...Lighthouse REMOTE Liquid Particle Counter 4 20mA output Operating Manual 2 8 248083296 1 Rev 3...
Page 29: ...Get Started 248083296 1 Rev 3 3 9 3 Wire System Figure 3 7 Three Wire Communication Diagram...
Page 50: ...Lighthouse REMOTE Liquid Particle Counter 4 20mA output Operating Manual 5 4 248083296 1 Rev 3...
Page 54: ...Lighthouse REMOTE Liquid Particle Counter 4 20mA output Operating Manual B 2 248083296 1 Rev 3...
Page 59: ...00...
Page 60: ......