Technical Data
248083296-1 Rev 3
5-3
OUTPUT
•
Outputs counts on CH1 and CH2 lines in 4-20 mA output currents.
•
When counts are greater than the alarm threshold, after a "suppress"
# of consecutive alarm conditions, the Alarm Relay is turned on.
ALARM
If the CH1 or CH2 count exceeds the programmed ALARM threshold
then:
•
The red SERVICE indicator on the instrument blinks.
•
The EXTERNAL ALARM relay, described below, is set.
SERVICE ALERT
If the service/background light voltage is greater than the Service High,
or less than Service Low limits, then:
•
CH1 and CH2 outputs are set to 2mA.
•
The red SERVICE indicator on the instrument is turned on steady.
•
If STATUS = 4 mA ==> No Alarms and No Service Alerts.
EXTERNAL ALARM
An ALARM condition, as described above, also closes a relay inside
the instrument. Contacts of that relay are connected to the DATA
connector, pins 8 (-) and 9 (+). The user may use an external power
supply and a buzzer or light stick to provide local indication of an
alarm condition. The contacts are rated at 40VDC, 1A maximum.
Summary of Contents for RLPC 0.1mm
Page 1: ......
Page 2: ...00...
Page 3: ...Lighthouse Worldwide Solutions REMOTE Liquid Particle Counters 4 20mA Output Operating Manual...
Page 10: ...Lighthouse REMOTE Liquid Particle Counter 4 20mA output Operating Manual ii 248083296 1 Rev 3...
Page 20: ...Lighthouse REMOTE Liquid Particle Counter 4 20mA output Operating Manual 2 8 248083296 1 Rev 3...
Page 29: ...Get Started 248083296 1 Rev 3 3 9 3 Wire System Figure 3 7 Three Wire Communication Diagram...
Page 50: ...Lighthouse REMOTE Liquid Particle Counter 4 20mA output Operating Manual 5 4 248083296 1 Rev 3...
Page 54: ...Lighthouse REMOTE Liquid Particle Counter 4 20mA output Operating Manual B 2 248083296 1 Rev 3...
Page 59: ...00...
Page 60: ......