Programming
248083296-1 Rev 3
4-3
1000:0100 ==> 6 seconds, 10,000 counts
1000:0010 ==> 6 seconds, 100,000 counts
0100:1000 ==> 60 seconds, 1000 counts
0100:0100 ==> 60 seconds, 10,000 counts
0100:0010 ==> 60 seconds, 100,000 counts
0010:1000 ==> 600 seconds, 1000 counts
0010:0100 ==> 600 seconds, 10,000 counts
0010:0010 ==> 600 seconds, 100,000 counts
All other combinations will be set to 300 seconds, 100,000 counts
Procedure to Set DIP Switches
1. Remove power from the instrument.
2. Set the DIP Switches to the desired Time and Range, using the
information above.
3. Apply power to the instrument.
Example Startup Echo, Reading DIP Switches:
Lighthouse Remote Instrument (4-20mA):
- reading dip switch:
Sample Time = 60 secs
Range1= 10000.
Range2= 1000.
Alarm Threshold = 750 counts
Suppress Alarms = 2
Service High = 3500 mV
Service Low = 125 mV