Programming
248083296-1 Rev 3
4-13
Type:
<ab1000>
Response:
<ab1000>Alarm Threshold= 1000 counts
Hyperterminal Alarm Alert without Alarm Suppression:
Note:
The field bk= 429
in this example is the
"backlight" function which
is a measurement of
scattered light in the
instrument. See
SET_SERVICE_HIGH and
SET_SERVICE_LOW
In this example, the Alarm Suppress is disabled (0).
secs=1: ch1-2: 285, 48 , bk= 429
secs=2: ch1-2: 487, 89 , bk= 429
secs=3: ch1-2: 1699, 1033 , bk= 1049
secs=4: ch1-2: 2572, 1320 , bk= 434
secs=5: ch1-2: 3337, 1575 , bk= 429
Sample: ch1-2: 3337, 1575 Range1= 10000, Range2= 5000.
Alarms: Threshold= 1000, Channel= 1, # in a row=1
secs=1: ch1-2: 615, 228 , bk= 434, alarm alert.
secs=2: ch1-2: 1115, 382 , bk= 429, alarm alert.
secs=3: ch1-2: 1595, 502 , bk= 429, alarm alert.
secs=4: ch1-2: 2029, 637 , bk= 429, alarm alert.
secs=5: ch1-2: 2381, 738 , bk= 429, alarm alert.
Sample: ch1-2: 2381, 738 Range1= 10000, Range2= 5000.
Alarms: Threshold= 1000, Channel= 1, # in a row=2
SET ALARM SUPPRESSION
<ae#>
Sets the alarm suppression for the configured alarm channel. The
number of alarm conditions have to be greater than # in order to turn on
the Alarm Relay.
For example, if Alarm Suppress is set to 2, then the device won’t report
an alarm condition via the external alarm output or the flashing service
light until the 3rd consecutive alarm condition.
Type:
<ae2>
Response:
<ae2>Suppress Alarms = 2
Summary of Contents for RLPC 0.1mm
Page 1: ......
Page 2: ...00...
Page 3: ...Lighthouse Worldwide Solutions REMOTE Liquid Particle Counters 4 20mA Output Operating Manual...
Page 10: ...Lighthouse REMOTE Liquid Particle Counter 4 20mA output Operating Manual ii 248083296 1 Rev 3...
Page 20: ...Lighthouse REMOTE Liquid Particle Counter 4 20mA output Operating Manual 2 8 248083296 1 Rev 3...
Page 29: ...Get Started 248083296 1 Rev 3 3 9 3 Wire System Figure 3 7 Three Wire Communication Diagram...
Page 50: ...Lighthouse REMOTE Liquid Particle Counter 4 20mA output Operating Manual 5 4 248083296 1 Rev 3...
Page 54: ...Lighthouse REMOTE Liquid Particle Counter 4 20mA output Operating Manual B 2 248083296 1 Rev 3...
Page 59: ...00...
Page 60: ......