Programming
248083296-1 Rev 3
4-15
Hyperterminal Service Alert Example:
secs=1: ch1-2: 667, 214 , bk= 463
secs=2: ch1-2: 2262, 1395 , bk= 449
secs=3: ch1-2: 2890, 1608 , bk= 434
secs=4: ch1-2: 4719, 2983 , bk= 4995, service alert.
secs=5: ch1-2: 4719, 2983 , bk= 4995, service alert.
Sample: ch1-2: 4719, 2983 Range1= 10000, Range2= 5000.
Alarms: Threshold= 1000, Channel= 1, # in a row=1
Service Menu
The Service Menu is available by typing
<ms>
.
Type:
<ms>
Response:
<ms>
*** Service Menu ***
<ia> Don't show 1 sec data
<ib> Show 1 sec data
<gb> Get Bkgnd
DON’T SHOW 1 SEC DATA
<ia>
Turns OFF 1 second data display.
Type:
<ia>
Response:
<ia>Don't show 1 sec data.
After starting the counter with <sa>, the device will wait until the
sample time interval passes before displaying the first sample.
SHOW 1 SEC DATA
<ib>
Turns ON 1 second data display.