background image

5

2

Attach sensor wires to electrodes

1.  Remove four electrodes from the electrode pouch.
2.  Fold over the open end of the pouch to protect the unused electrodes 

from drying out.

Note:

 Electrodes should be stored at room temperature.

 

3.  Snap an electrode onto each of the 

      four lead wires that are attached to                                                                     

      the sensor.

4.  Using the peel tab, peel each electrode                                                     

from the backing and discard the backing.

4

Place electrodes on your body

4

1.  Place each color coded electrode on healthy, non-irritated skin according 

to the diagram on the previous page (pg.1). 

Do not apply electrodes to skin that has an open cut or is already irritated.    

If skin is cut or irritated, apply the electrode in close proximity to the required 

location on clean, non-irritated skin.

2.  Apply each electrode by gently pressing down on the center and moving 

toward the edge in a circular motion to assure firm attachment to the 

skin.

Place the white lead wire with electrode 3 fingers width below the         

midpoint of the collarbone.
Place the red lead wire with electrode above your last rib on the left side 

aligned with the midpoint of the collarbone.
Place the black lead wire with electrode on your left side, about 4-5 inches 

horizontally to the left of the red electrode, in line with the center of your 

armpit.
Place the green lead wire with electrode on the left side of the abdomen. 

Note:

 If necessary, lift the left breast to get the proper electrode placement 

of the red and black electrodes.

How to record a Manual Event

You can record a manual event if you experience dizziness, chest pain, or 

other symptom(s). 

Initiate recording with sensor or Monitor.

Monitor Layout

Press the sensor side buttons simultaneously and release 

the buttons when there is a beep. 

OR

Select “Record Event” at the bottom of Monitor homescreen.  

Select “Yes” to send recording, or “No” to cancel.

Choose your symptoms and activities on Monitor Screen

1.  Choose the symptom(s) you felt, or select “No Symptoms”.

2.  Select “Next” at the bottom of the screen.

3.  Choose the activity you were engaged in.

4.  Select “Send” to send the manual event.

3

4

On/Off Button

Monitor Charge Port

Battery Indicator shows the level of battery 

charge and if the Monitor is in charge mode.

Emergency Call Button 

(call can be canceled within 10 sec.)

Icon alerts may appear on Monitor homescreen. 

Alerts will disappear once issue has been resolved.

Summary of Contents for MCT 3L

Page 1: ...Return D Technical Alarms End of Service Date If not provided here call LifeWatch at 0850 433 25 88 Monitor LifeWatch Mobile Cardiac Telemetry 3 Lead MCT 3L A Sensor with Wires Electrodes INTERNATIONAL PART D Technical Alarms Monitor Battery Select OK and charge the monitor Sensor Battery Replace sensor battery and select OK to continue Note The sensor alerts its low battery status with one long b...

Page 2: ...may remove the electrodes in the shower by using a wet warm washcloth If you remove the electrodes in the shower please remove lead wires beforehand 2 Clean your skin Gently clean away excess adhesive or dead skin between electrode changes using a wet warm washcloth Allow skin to thoroughly dry How to Return MCT 3L Box to LifeWatch at End of Service 1 Place ALL MCT 3L components into the original ...

Page 3: ...igned with the midpoint of the collarbone Place the black lead wire with electrode on your left side about 4 5 inches horizontally to the left of the red electrode in line with the center of your armpit Place the green lead wire with electrode on the left side of the abdomen Note If necessary lift the left breast to get the proper electrode placement of the red and black electrodes How to record a...

Page 4: ...continue A LifeWatch client care representative will carefully walk you through each step of the process to get you started STOP 5 6 7 Continuous Heart Monitoring The sensor continuously records your heart rhythm which is transmitted to the Monitor for analysis Any irregular rhythm is immediately sent to a LifeWatch clinical technician for review DO Keep the Monitor within 3 meters of the worn sen...

Reviews: