51335-006
EEG System
1.4
The system and its parts
The R40 EEG System is a multi-channel electroencephalograph designed for use in routine EEG and lab
monitoring applications.
The system uses either the R40 EEG amplifier or the R40 (24) EEG amplifier.
The R40 EEG amplifier is a compact USB 40-channel amplifier which incorporates 32 referential channels and 8
polygraphic channels with built-in calibration and electrode impedance measurement. A Nonin XPOD interface is
provided, a Patient Event input, 2 Aux DC inputs and an Electrocap connector.
The R40 (24) EEG Amplifier is a reduced channel variant of the R40 amplifier and incorporates 24 referential
channels and 4 polygraphic channels. In all other respects the R40 (24) Amplifier is identical to the R40.
The system includes a PC for control and storage, a video monitor, a medical-grade power supply and an optional
USB Photic Stimulator and arm. The system is mounted on a special wheeled cart, which houses all components
and allows convenient mobility.
The system uses a specific PC and medical grade power supply. Refer section 3.1 for details
The PC or laptop must only be powered using the medical-grade mains power supply, as supplied
or authorised by Lifelines.
The R40 EEG System comprises the following components:
Component
Part Number
R40 Amplifier
1326
R40 (24) Amplifier
1411
Amplifier USB Cable
1277
Cart, adjustable height
1296
Arm (for Photic or R40 Amplifier)
1291
Photic Stimulator
1290
Photic USB Cable
1241
Xpod Pulse Oximeter Nonin
1327
Patient Event Pushbutton
1353
NOTE:
The Oximeter sensor is a consumable and is not supplied by Lifelines. Only use the ‘PureLight’ sensors
specified by Nonin for use with their Oximeters.
NOTE:
If using the Photic Stimulator refer to the User Manual supplied with it.